Short notes

Biomedical Instrumentation Short Notes

A disposable Ag/AgCl electrode stuck on the chest of a patient in an ICU picks up a 1 mV peak ECG signal riding on top of a 1.5 V half-cell potential and 50 Hz mains interference that can be 1000× larger than the signal — the INA128 bio-amplifier must reject the 50 Hz common-mode noise by more than 80 dB while amplifying the differential ECG by 60 dB. Biomedical instrumentation design is fundamentally the problem of extracting microvolt-to-millivolt physiological signals from a hostile electrical environment while ensuring the patient is never at risk of even microamp leakage currents.

EI

How it works

The ECG (electrocardiogram) signal is 0.5–4 mV peak, bandwidth 0.05–150 Hz; EMG (electromyogram) is 50 μV–5 mV, 20 Hz–2 kHz; EEG (electroencephalogram) is 5–300 μV, 0.5–100 Hz. Ag/AgCl electrodes are preferred because their half-cell potential is stable at +0.222 V, minimising offset at the amplifier input. A bio-amplifier (cardiac) must have: input impedance > 10 MΩ, CMRR > 80 dB, gain ≈ 1000, noise < 1 μVrms referred to input, and leakage current < 10 μA (IEC 60601-1 safety limit). Patient isolation is achieved using an isolation amplifier (ISO124) or optical coupling, keeping the patient circuit floating from mains earth — a fault current of > 50 μA through the heart can induce ventricular fibrillation. Pacemakers deliver 1–10 V pulses of 0.5–2 ms duration at 60–100 ppm.

Key points to remember

The 50 Hz interference on ECG is a power-line common-mode signal rejected by the CMRR of the differential amplifier — improving electrode contact quality (reducing skin-electrode impedance below 10 kΩ) also reduces the converted interference at the amplifier input. ECG lead configuration: Lead I = VLA − VRA, Lead II = VLL − VRA, Lead III = VLL − VLA; these obey Einthoven's triangle: Lead I + Lead III = Lead II. IEC 60601-1 specifies that patient leakage current must not exceed 10 μA under normal conditions and 50 μA in single-fault conditions. The notch filter (50 Hz, Q ≈ 30) using a twin-T network removes power-line interference without distorting the ECG P and T waves significantly. Cardiac output monitors, pulse oximeters (SpO₂), and blood pressure transducers also fall under biomedical instrumentation and may appear in exam questions alongside ECG.

Exam tip

The examiner always asks you to list the specifications of a bio-amplifier for ECG measurement — always include input impedance (> 10 MΩ), CMRR (> 80 dB), gain, bandwidth (0.05–150 Hz), leakage current limit (< 10 μA), and the reason for patient isolation in your answer.

More Measurements and Instrumentation notes